Pulser8KApp module Release Notes Version 1.0.22 (Jun-6-2024) - Fixed: Do not restore VBias after a test is done. Restoring a VBias may change a device state in SOT devices. Version 1.0.21 (Apr-24-2024) - Added: SOT support is added. - Fixed: VSwitch does not generate correct set pulse if "Reset pulse before regular" is enabled. Version 1.0.20 (Jan-17-2024) - Added: In Prob-Error Sweep test, threshold results for high and low states are added. - Added: In Prob-Error Sweep test, "Read Out Delay" parameter is calculated automatically and reported as a new result. - Added: In Prob-Error Sweep test, lookup of comparator threshold for AC mode is modified. Version 1.0.19 (Dec-14-2023) - Fixed: Prob-Error Sweep test does not generate correct reset pulses if test is sweeping on "Negative Side". Version 1.0.18 (Sept-6-2023) - Added: reset pulse option in HTransfer test. Version 1.0.17 (Aug-23-2023) - Fixed: Release for a new STTPulserLib3 v. 1.5.0 Version 1.0.16 (June-30-2023) - Fixed: Remove "Keep" choice in "Auto-Detect By" parameter in Prob Error test. Version 1.0.15 (June-19-2023) - Fixed: The min/max read VBias is changed to 3V in tests - Added: reset field parameters are added in MTJ test Version 1.0.14 (May-31-2023) - Fixed: The max read VBias is changed to 3V in VSwitch test Version 1.0.13 (May-1-2023) - Fixed: In H-Transfer test, change auto QST gain adjustment criterion to fix problem in some DUT. - Fixed: In V-Transfer test, add 20 ms delay after reset pulse so that it has enough time to reset DUT. Version 1.0.12 (Apr-13-2023) - Fixed: If reset pulse width is larger than 655 us in Prob-Error Sweep test, runtime error will be displayed. Version 1.0.11 (Apr-10-2023) - Fixed: The reset pulse width limit in Prob-Error Sweep test is fixed to match hardware limit (5 ms max) - Fixed: wrong statistical results if "Cycles" > 1 in H-Transfer test Version 1.0.10 (Mar-23-2023) - Fixed: Low Pos Rh and Rl results in V-Transfer are swapped - Fixed: pos and neg low bias results are swapped in V-Transfer test if sweeping from pos bias to neg bias Version 1.0.9 (Mar-21-2023) - Fixed: "Auto QST Gain" gives wrong QST gain when negative VBias is used in H-Transfer test - Added: new results in V-Transfer test Version 1.0.8 (Mar-17-2003) - Added: "Auto QST Gain" parameter is added in H-Transfer test. - Added: "Auto Pulser Gain" parameter is added for tests using Pulser gain. - Added: "Overflow" and "FinalPulserGain" results are added in tests: H-Transfer, Read Disturb, RVHScan, Single Pulse, V Break Down, V Switch, V Transfer, Write Prob Version 1.0.6 (Mar-03-2023) - Fixed: MTJ resistance test gives wrong "overflow" result if negative bias is used. - Fixed: H-Transfer test uses QST to measure reference resistance Version 1.0.5 (Feb-28-2023) - Fixed: wrong unit for Threshold result in Error Rate test - Fixed: H-Transfer test does not work if "Amplitude Gain" is selected Version 1.0.4 (Feb-24-2023) - Fixed: "Pulse Width (us)" was used wrongly to reset when Asymmetric Pulse is enabled. Version 1.0.3 (Feb-21-2023) - Added: Pulser gain selection to tests - Added: All tests, except HTransfer, use Pulser for measurement - Fixed: "Asymmetric Pulses" in Prob-Error Sweep is never enabled - Fixed: When "Enable Stop" is enabled, Endurance test stops immediately even test results are good. Version 1.0.2 (Jan-24-2023) - Added: adjustment to measured resistances Version 1.0.1 (Jan-20-2023) - Added: make use of pulser-QST VBias Version 1.0.0 (Jan-12-2023) - Initial Release