top of page

FMRA.NET Test Module v4.0.5B2

FMRA.NET Test Module v4.0.5B2

FMApp Release Notes

Version 4.0.5 (June 27 2016)
-Fixed: FMRTransfer: Was using fixed default results used in debugging
-Fixed: FMRSweep: subset plotting color error

Version 4.0.4 (June 27 2016)
-Fixed: All Tests had Frequency and Amplitude Axis labelling inverted
-Fixed: RPLStability Test: Did not show correctly What to plot in plot selection
-Fixed: FMRTransfer: DBNull error in autoplotting of plots
-Fixed: FMRSweep: Did not show correctly What to plot in plot selection when form initially opened

Version 4.0.3 (Sep 25 2015)
-Fixed: compatibility to Quas97 7.31.14

Version 4.0.2 (Apr 20 2015)
-Fixed: RPL Test DBNull error while running
-Fixed: RPL Test subset labels displayed along with annotations
-Fixed: wrong Y Axis label in the FMR Sweep test
-Fixed: wrong Y Axis label in the RPL Stability test
-Fixed: SFT Test column already exist if 0oe was included in the sweep

Version 4.0.1 (Apr 17 2015)
.NET version (minimum Quasi97 v7.21.4 is required)

Version 3.2.7 (Apr 16 2014)
-Fixed: External spectrum filter results were not being saved. When setupfile closed, reverted to default values

Version 3.2.6 (Apr 16 2014)
-Fixed: fix the runtime error when closing quasi97
-Fixed: Previous release could not run on windows 7, due to licensing error

Version 3.2.5 (Apr 16 2014)
-Fixed: Removed MSForms - to fix the runtime error when closing quasi97

Version 3.2.4 (Feb 03 2014)
-Fixed: Allow setting of Increment below ISi internal mimimum, when system is external spectrum capable
-Fixed: Set points on external spectrum analyzer to match {From(GHz,To(GHz), Increment} points needed.
-Added: Capability to return raw data from spectrum analyzer without any ISI processing. Use by adding fmr.SkipProcessing into stress menu

Version 3.2.3 (DEC 03 2013)
-Fixed: allow setting of frequencies below 200MHz and above 13Ghz when system is external spectrum capable

Version 3.2.2
-Fixed: When setting freqresponse got array error. introduced in 3.2.0

Version 3.2.1
-Fixed: When setting external analyzer, object error occured. introduced in 3.2.0

Version 3.2.0 (Oct 8 2013)
- Added: ISI2010Lib .NET compatibility

Version 3.1.7 (Aug 27 2013)
- Fixed: In FMRSweep test if sweeping static field, did not restore static field back to 0.
- Added: Substituted Field Angle sweep item with QPS Splitter.Field Angle XY sweep item.

Version 3.1.6 (Aug 15 2013)
- Fixed: Error in FMRApp Terminate, if isi2010 not present
- Fixed: Quasi-starter memory could not be read error when closing Quasi97

Version 3.1.5 (Aug 15 2013)
- Fixed: License form displayed with error in V7
- Fixed: multiple error if FMR hardware not detected
- Fixed: Stiffness field test field_to parameter was not displayed correctly on the form

Version 3.1.4 8-01-2013
- Fixed: If license problem, showed error instead of showing license form

Version 3.1.3 Jun-28-2013
- Fixed: When running FMR Sweep, if second sweep value not selected, gave object error

Version 3.1.2 Jun-28-2013
- Added: Compensate for MR Resistance

Version 3.1.1 Jun-19-2013
- Added: Detects which preamp connected and sets to internal or external on startup.
- Fixed: When used with vb.net fmrsettings form could not be shown.

Version 3.1.0 Jun-13-2013
- Fixed: Memory leak caused by proE3 (migrated to ProE7)

Version 3.0.2 Jun-01-2013
- Added: Modified all tests to work with stress option capability for changing spectrum selected.

Version 3.0.1 May-29-2013
- Fixed: Filter form not showing
- Added: SpectrumAnalyzer compatibility
- Added: Licensing validation for either spectrumanalyzer or FMRApp

Version 3.0.0 May-23-2013
- Added: Compatible to Quasi97 .NET

Version 2.2.5 May-23-2013
- Added: Fixed Error with FMRSweepTest capability to log raw data

Version 2.2.4 Apr-12-2013
- Added: FMRSweepTest capability to log raw data

Version 2.2.3 Mar-27-2013
- Added: display device row and col number of device tested on all plots

Version 2.2.2 Mar-18-2013
- Added: dsiplayed device row and col number of device tested on some plots


Version 2.2.1 Jan-23-2013
- Fixed: If preconditioning used in FMRSweep, RPL Stability or FMR Transfer test, was not restored back to initial values
- Fixed: Division by Zero error in SFT test

Version 2.2.0 Jul 17 2012
- Added: supports non-admin mode

Version 2.1.4 Mar 16 2012
- Added: Fine sweep at max peak
- Added: Skip subtract bias off value
- Added: Subtract system noise option

Version 2.1.1 Mar 14 2012
- Fixed: In FMR Sweep test, When trying to sweep external stress item, generated an error "RunExternalStress
item in the wrong format".

Version 2.1.0 Feb 23 2012
- Added: Option to automatically add FMR and QPS Splitter to the hardware configuration

Version 2.0.3 Feb 20 2012
- Fixed: Did you save "Subtract baseline noise" setting

Version 2.0.2 Feb 06 2012
- Fixed: Errors if FMR not defined in the hardware configuration file

Version 2.0.0 Jan 30 2012
- Added: Compatilibity to ISI2010Lib and QPSSplitLib, operating through Quasi97 Hardware Options.

Version 1.3.1 May 16 2011
- Fixed: Error in Filter setup form: If subtract noise checked. Form loaded with error message.

Version 1.3.0 May 6 2011
- Added: Capability to change Amplitude units between db, nv, nv^2.

Version 1.2.7 Sep 2 2010
- Added: Max Frequency SW1 and SW2 results to the FMR Sweep Test.

Version 1.2.6 Jul 15 2010
- Fixed: The peak was not interpolated, which resulted in >50MHz digitization.
- Added: Programmable AC Gain parameter for QST tests

Version 1.2.5 Nov 19 2009
- Fixed: Field parameter in FMR Sweep test was not taken into account
- Fixed: In all tests, field polarity was not taken into account, which resulted in mismatch of the field in transfer curve and other quasi tests vs FMR tests.

Version 1.2.4 Oct 22 2009
- Fixed: RPLT Stability test showed an error "CopyRawData: Subscript out of range", when running in production and logging raw data, and number of cycles was more than 1

Version 1.2.3 Oct 8 2009
- Fixed: FMRSweep test, Min sw1 and Min sw2 reported the location of minimum amplitude, not frequency.

Version 1.2.2 Sep 18 2009
- Fixed: Abort function in RPLT Stability
- Fixed: Abort function in FMR Transfer test
- Added: Null Data Gaps in the FMR Angle Sweep Test

Version 1.2.1 Sep 8 2009
- Fixed: Did not plot results someimes if the test was aborted

Version 1.2.0 Sep 2 2009
- Added: Capacitance result for reference to all tests
- Added: Marker and integral noise results to all of the tests

Version 1.1.7 Aug 20 2009
- Fixed: Removed parabola fitting to find the center
- Fixed: Changed PW50 calculation to include only 3 data points (+/-1 around found value).
- Added: Marker and Integral Noise results to the FMRS and FMR Sweep tests
- Added: Peak Detect Overlay function
- Added: Threshold for peak area magnitude. Peaks with lower areas are discarded.
- Added: FMR Settings are now saved to the MDS setup file
- Added: Capacitance result for reference, if scaling option is enabled

Version 1.1.6 Aug 05 2009
- Fixed: Scaling option did not get saved

Version 1.1.5 Aug 05 2009
- Fixed: FMR Angle Sweep test did not display Frequency or Amplitude plots if two parameters were swept.
- Fixed: Adaptive parameters did not work in any of the tests
- Added: Scaling option for the FMR results
- Fixed: The colors on the FMR transfer curve test were not visible

Version 1.1.4 July 30 2009
- Added: icons to the tests

Version 1.1.0 July 23 2009
Initial Release

Version 1.0.5 June 9 2009
Internal Release
Fixed: in FMRA test all plots where shown instead of the last one

Version 1.0.4 Sep 30 2008
Internal Release
Added: FMR Field Angle Sweep Test

Version 1.0.2 Sep 10 2008
Internal Release
Fixed: Amplitude reported in uv, should be dB

Version 1.0.1 Sep 10 2008
Internal Release
Fixed: did not read the specturm analyzer type correctly from the ini file

Version 1.0.0
Initial Release

bottom of page